  • Ensure the load cell is mounted securely using the specified torque for each capacity.
  • Connect the shielded cable according to the color function provided.
  • Apply the force to be measured to the active end of the load cell.
  • Avoid exceeding the overload protection limit of the load cell.

KEMELO LIEKETSENG

Interface-SSM-50-1.1-Load-Cell-fig-1

TŠEBELETSO MOTLATSI

The SMT1 through SMT2 Model load cells are supplied with a 4-conductor shielded cable (AWG28) 5ft (1.5m) long.

 

Interface-SSM-50-1.1-Load-Cell-fig-2

Khoutu ea Wiring e Lumellana le ISA S37.8 'Specification and Test for Strain gage Force Transducers' le SMA Load Cell Terminology.

KOPO LINTLHA

  • MOUNTING TORQUE SPECIFICATIONS
    At least one diameter thread engagement is desirable for mounting the load cell. Jam nuts may be used, but are not required. React torque only on the load cell structure adjacent to the mounting surface or jam nut, and limit torque to the following recommended specifications.
MAKHONO 1.1lbf, 5N 2.2lbf, 10N 11lbf, 50N 22lbf, 100N 112lbf, 500N 225lbf, 1000N 11lbf, 50N
inch-lb 5 5 20 20 100 200 300
Nm 0.6 0.6 2 2 11 22 33
  • The SMT Model load cells incorporate a unique overload protection system that provides up to 10 times the rated load in both tension and compression without degradation of performance. This feature makes SMTs very useful in process, medical, and pharmaceutical weighing, as well as automotive, aircraft, dynamometer, and small rocket engine test applications where forces need to be measured under comparatively large overload conditions.
  • The force to be measured should be applied to the active end of the cell to eliminate possible errors due to cable interaction. The above figure can be utilized to identify the active end of the load cell.

LIEKETSENG MOSEBETSI

  • Nominal Output – mV/V  2
  • Input Resistance Ohms  350 + 50 / -3.5
  • Output Resistance Ohms  350 +/- 3.5
  • Recommended Excitation Vdc  10
  • Nonlinearity – %RO  < +/- 0.05
  • Hysteresis – %RO    < +/- 0.03
  • Temp. Range Compensated  -15 to 50°C
  • Temperature effect on zero%-RO %RO/°C    < +/- 0.15
  • Zero Balance – %RO    < +/- 3.0

TS'ELISO

LBH

  • Q: Can the load cell be used in medical weighing applications?
  • A: Yes, the load cell is suitable for medical weighing applications due to its unique overload protection system.
  • Q: What is the recommended mounting torque for the load cell?
  • A: The recommended mounting torque varies based on the capacity of the load cell. Refer to the mounting torque specifications provided in the manual.