Selelekela
The Detecto 6800-AC Portable Bariatric Scale is designed for clinical weighing applications, offering a low-profile design for patient comfort and ease of transport. This manual provides essential information for the proper setup, operation, and maintenance of your scale.
Key features include a 1,000 lb / 450 kg capacity, BMI calculation, EMR/EHR readiness, an integral carrying handle, and a low-profile design for easy patient access.
Sehlahisoa se Felletseview

This image shows the Detecto 6800-AC portable bariatric scale. It features a large, square, low-profile weighing platform with a textured, skid-proof rubber surface. A metal carrying handle is visible on the side of the platform. The digital indicator, connected by a cable, displays '75.0' and has buttons for 'ZERO', 'UNITS', 'LOCK/RELEASE', 'ON/OFF', and 'BMI/ENTER'.
Litaelo tsa ho seta
- Tlosa Sekala: Carefully remove all components from the packaging. Verify that all parts are present: the weighing platform, the model 750 weight display, and the 10 ft (3m) indicator cable.
- Position the Platform: Place the 24-inch (61 cm) square weighing platform on a firm, level surface. Ensure the surface is stable and free from vibrations to guarantee accurate readings.
- Connect the Indicator: Connect the 10 ft (3m) indicator cable from the weighing platform to the model 750 weight display. Ensure the connection is secure.
- Phepelo ea motlakase:
- Sesebelisoa sa AC: Connect the provided AC adaptor to the indicator and then to a suitable power outlet.
- Matla a Battery: For portable use, the scale can be powered by 6 AA batteries (not included). Open the battery compartment on the back of the indicator and insert the batteries, observing correct polarity.
- Motlakase oa Pele o Buletsoe: Tobetsa ea BONAHALA/TSHA button on the indicator to power on the scale. Allow a few seconds for the display to stabilize and show "0.0".
Litaelo tsa Ts'ebetso
Basic Weighing Procedure
- Matlafatsa: Tobetsa ea BONAHALA/TSHA button. Wait for the display to show "0.0".
- Ho fokotsa Sekala: If the display does not show "0.0" when empty, press the ZERO konopo ho e seta bocha.
- Kgetha Diyuniti: Tobetsa ea LIEKETSENG button to toggle between pounds (lb) and kilograms (kg). The selected unit will be displayed.
- Boima: Carefully assist the patient onto the center of the weighing platform. Ensure the patient remains still. The weight will be displayed on the indicator.
- Lock/Release: The LOCK/RELEASE button can be used to hold the displayed weight after the patient steps off, or to release the held weight.
- Tlosa motlakase: Tobetsa ea BONAHALA/TSHA button again to turn off the scale when not in use.
Palo ea BMI
The Detecto 6800-AC scale can calculate Body Mass Index (BMI).
- Weigh Patient: Follow the basic weighing procedure to obtain the patient's weight.
- Enter Height: Once the weight is stable, press the BMI/ENTER button. The display will prompt for height input. Use the arrow keys (PHAHAMA/TLASE) to adjust the height value.
- Confirm Height: Tobetsa BMI/ENTER again to confirm the height. The scale will then display the calculated BMI.
Tlhokomelo
- Ho hloekisa: Clean the weighing platform and indicator regularly with a damp cloth and mild, non-abrasive cleaner. Do not use harsh chemicals or immerse the unit in water. Ensure the textured rubber surface is clean to maintain its skid-proof properties.
- Phetoho ea Betri: If using battery power, replace all 6 AA batteries when the low battery indicator appears on the display or if the scale fails to power on. Always replace with new batteries of the same type.
- Polokelo: When not in use for extended periods, store the scale in a dry, temperate environment. The integral carrying handle facilitates easy transport and storage.
- Tlhahlobo: Periodically inspect the indicator cable for any signs of damage or wear. Ensure all connections are secure.
Ho batle phoso
| Bothata | Lebaka le ka khonehang | Tharollo |
|---|---|---|
| Sekala ha se sebetse. | No power from AC adaptor; dead batteries; loose cable connection. | Check AC adaptor connection and power outlet. Replace batteries. Ensure indicator cable is securely connected. |
| Litemana tse sa nepahalang. | Scale not on a level surface; patient movement; external vibrations; debris under platform. | Relocate scale to a firm, level surface. Ensure patient remains still. Remove any obstructions under the platform. |
| Display shows "Err" or other error codes. | Overload; internal malfunction. | Ensure weight does not exceed 1000 lb / 450 kg capacity. Power off and on again. If error persists, contact customer support. |
| Cannot change units. | Unit button not pressed correctly; scale in a locked state. | Tobetsa ea LIEKETSENG button firmly. Ensure the scale is not in a locked mode (press LOCK/RELEASE ha ho hlokahala). |
Litlhaloso
- Mohlala: 6800-AC
- Bokhoni: 1,000 lb x 0.2 lb / 450 kg x 0.1kg
- Litekanyo tsa Sethala: 24 inches x 24 inches (61 cm x 61 cm) square
- Lisebelisoa tsa Sethala: Skid-proof, textured rubber surface
- Boima ba Ntho: 45 Liponto
- Indicator Cable Length: 10 ft (3m)
- Matla: AC Adaptor (included) or 6 AA batteries (not included)
- Likaroloana: BMI Calculation, EMR/EHR Ready, Integral Carrying Handle
- Tumellano: OIML EC Class III model available (specific model may vary)
Warranty le Tšehetso
For detailed warranty information, please refer to the warranty card included with your product or visit the official Detecto websebaka. Lipehelo le maemo a tiisetso li ka fapana.
If you require technical assistance, parts, or service, please contact Detecto customer support. Contact information can typically be found on the manufacturer's websetsheng kapa sephuthelong sa sehlahiswa.
It is recommended to register your product upon purchase to facilitate warranty claims and receive important product updates.